Long-term complications evidenced studying the explanted Gore® HELEX® atrial septal defect occluder seven years after implantation: a case report

Main Article Content

Jurijs Sekretarjovs
Mārtiņš Kalējs
Ainārs Rudzītis
Ivars Brečs
Hermanis Sorokins
Sandra Skuja
Pēteris Stradiņš
Valērija Groma

Abstract

We present the results of the first morphological study of a Gore® HELEX® Septal Occluder 30 mm that was explanted seven years after interventional implantation due to a significant left-to-right shunt (7 mm) which resulted from the stretching of the concomitant patent foramen ovale by the occluder after atrial septal defect closure. Complete endothelialization of the surface of the device, the formation of the connective tissue around the implant, minor chronic inflammation, the appearance of foreign body giant cells and weakened myocardial cells adjacent to the implant as well as enhanced expression of matrix metalloproteinases were demonstrated.

Article Details

How to Cite
Sekretarjovs, J., Kalējs, M., Rudzītis, A., Brečs, I., Sorokins, H., Skuja, S., Stradiņš, P., & Groma, V. (2017). Long-term complications evidenced studying the explanted Gore® HELEX® atrial septal defect occluder seven years after implantation: a case report. Papers on Anthropology, 26(2), 105-112. https://doi.org/10.12697/poa.2017.26.2.11